This function allows for reading in LS Pro data based on their text export of signals. Signals can be exported directly from the LS Pro system.
The LabSystem Pro was acquired by Boston Scientific from the original company Bard. They are a common electrophysiology signal processing device for visualization and measurement of intracardiac signals.
Value
An egm
class object that is a list of eps
signals the format of a
data.table
, with an attached header attribute that contains
additional recording data.
Data Export
The steps to data export are as follows.
Start LabSystem PRO
Open a patient record
Display a waveform recording in a Review Window
Scroll to a point of interest in a waveform recording
Right click on the review window to the left of the region of interest
Select an Export option, either a default time range or the entire visible page (which depends on the sweep speed).
Data Format
[Header] Recording info – contains (example):
[Header]<CR><LF>
File Type: 1<CR><LF>
Version: 1<CR><LF>
Channels exported: 22<CR><LF>
Samples per channel: 5000<CR><LF>
Start time: 6:55:24<CR><LF>
End time: 6:55:29<CR><LF>
Ch. Info. Pointer: 320<CR><LF>
Stamp Data: T<CR><LF>
Mux format: 0<CR><LF>
Mux Block Size: <CR><LF>
Data Format 1<CR><LF>
Sample Rate: 1000Hz<CR><LF>
[Header] Channel info (per channel example):
Channel #: 1<CR><LF>
Label: III<CR><LF>
Range: 5mv <CR><LF>
Low: 1Hz<CR><LF>
High: 100Hz<CR><LF>
Sample rate: 1000Hz<CR><LF>
Color: 0000FF<CR><LF>
Scale: -7<CR><LF>
[Data] As described below:
-256,-1056,576,-256,320,-736,144,576,-592,176,608,240,176,-560,496,-
144,0,0,-32,-48,-32,-80<CR><LF>
Channel Data is interleaved in the example above (sample indexed at 1):
1 | 2 | 3 | ... | 22 |
Ch1:1 | Ch2:1 | Ch3:1 | ... | Ch22:1 |
Ch1:2 | Ch2:2 | Ch3:2 | ... | Ch22:2 |
Ch1:3 | Ch2:3 | Ch3:3 | ... | Ch22:3 |
... | ... | ... | ... | ... |
Ch1:5000 | Ch2:5000 | Ch3:5000 | ... | Ch22:5000 |